The CMS8000VET Veterinary Monitor has abundant monitoring functions and can be used for monitoring cats, dogs, and other animals. Users may select different parameter configurations according to different requirements. It adopts 100-240V~,50/60Hz for power supply, 12.1'' TFT color LCD for displaying real-time data and waveform, and up to 8-channel waveform and all monitoring parameters can be displayed simultaneously. In addition, it can be connected to the central monitoring system via a wired/wireless network so as to form a network monitoring system.
This device can monitor such parameters as ECG, RESP, NIBP, SpO2 and dual-channel TEMP, etc. It integrates parameter measurement modules, displays, and recorder in one device to form compact and portable equipment. At the same time, its built-in replaceable battery provides convenience for moving.
Functions
Standard parameters: ECG, RESP, SpO2, PR, NIBP, dual-channel TEMP
1)ECG Heart rate (HR)
ECG waveform
Arrhythmia and ST-segment analysis
2)RESP Respiration rate(RR)
Respiration waveform
3)SpO2 Oxygen saturation(SpO2)
Plethysmogram(PLETH) waveform
Pulse rate(PR)
Bar graph
4)NIBP Systolic pressure(SYS), Diastolic pressure(DIA), Mean Pressure(MEAN)
5)TEMP T1, T2, TD
It has abundant functions, such as audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation, etc
Features
1)12.1'' TFT color LCD, multi-language interface(Simplified Chinese, Traditional Chinese, English, French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian, Kazakhstan, Polish, Czech).
2)Fanless design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.
3)With the standard interface, OxyCRG, trend chard, big character interface, and view bed, convenient to observe.
4)Operation mode: keys and buttons.
5)Display up to 8-channel waveform on one screen.
6)Display 7-lead ECG waveform on one screen, and with ECG step function.
7)Adopt digital SpO2 technology, anti-motion, and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.
8)NIBP measurement mode: Manual/AUTO/STAT, storage for 4800-group NIBP data.
9)Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
10)Drug concentration calculation and titration table functions.
11)Print trend table data with one key.
12)Connect to Central Monitoring System by 3G, WiFi, or wired mode.
13)AC/DC, built-in rechargeable lithium battery achieve uninterrupted monitoring.
14)Anti-high frequency surgical unit, defibrillation-proof(special leads are necessary).
Performance
▲ECG
Lead mode: 3-lead or 5-lead
Lead selection: I, II, III, aVR, aVL, aVF, V
Waveform: 5-lead, 2-channel
3-lead, 1-channel
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s
HR:
Measurement and alarm range: 15~350bpm
Accuracy: ±1% or ±1bpm, whichever is greater
Alarm accuracy: ± 2 bpm
Resolution: 1 bpm
ST-segment monitoring:
Measurement and alarm range: -2.0mV~+2.0mV
Accuracy: -0.8mv~+0.8mv ±0.04mv or ±10%, whichever is greater
Other range: unspecified
Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,
PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Pacemaker: yes
▲RESPIRATION
Method: R-F(RA-LL) Impedance
Respiration rate:
Measurement and alarm range: 0~150rpm
Resolution: 1 rpm
Measurement accuracy: ±2 rpm
Alarm accuracy: ±3 rpm
Apnea alarm: 10~40s
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s
▲NIBP
Method: Oscillometry
Mode: Manual/AUTO/STAT
Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
Measurement period in STAT mode: 5 minutes
Measurement and alarm range: 10 ~ 270 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ±3 mmHg
Measurement accuracy:
Maximal mean error: ±5 mmHg
Maximal standard deviation: 8mmHg
Over-pressure protection:
Other modes: 297±3 mmHg
Dog mode: 240±3 mmHg
Cat mode: 147±3 mmHg
▲SpO2
Measurement and alarm range: 0 ~ 100%
Resolution: 1%
Measurement accuracy: 70%~100%: ±2%;
0%~69%: unspecified
▲PR
Measurement and alarm range: 30 ~ 250 bpm
Measurement accuracy: ±2 bpm or ±2%, whichever is greater
▲TEMP
Channel: dual-channel
Measurement and alarm range: 0 ~ 50℃
Resolution: 0.1 ℃
Accuracy: ±0.1 ℃
▲Power supply: 100-240V~,50/60Hz
▲Safety classification: Class I, type CF defibrillation-proof applied part